Image 16-6
Draining circuit, excluding light processor circuit
3. Fill the syringe with air. Press in the valve of the male fitting when extracting air.
Image 16-7
4. Couple the air filled syringe to the inlet of the cooling circuit and force air into the circuit via the syringe. Liquid will be expelled
from the circuit exit side into the empty bottle.
R59770072 FLM SERIES 19/03/2007
OUT
NO OUT
16. FLM liquid cooling circuit
161