OPERATION

2.7 Imbalance Detection

In case the rotor is not equally loaded, the drive will turn off, during acceleration. The rotor decelerates
to a standstill.
When in the display "time" (A-3), the word "error" (M11) along with the number "01" appears, the
weight difference of the samples are too large. Weigh out the samples exactly!
Load the rotor as described in Chapters: 2.1.2 and 2.1.3.
When inside the display "time" (A-3), the word "error" along with the number "02" appear, (see Figure
27)
Potential Reason for the Error Screen: The imbalance switch is defective.
Figure 27
21
Z326_V2.16
© Hermle Labortechnik GmbH