Table 5: Error Messages
Error-No.:
Description
1
Imbalance arose
2
Imbalance sensor is defective
4
Imbalance switch has been activated for longer than 5 seconds
8
Transponder in the rotor is defective
11
Temperature sensor is defective
12
Chamber over temperature
Leap of speed is too large between two measurements
14
CLOSE lid
33
Open lid while motor is running
34
Lid contact defective
38
Lid motor is blocked
40
Communication with frequency converter disturbed during start
41
Communication with frequency converter disturbed during stop
42
Short circuit in the frequency converter
43
Under-voltage frequency converter
44
Overvoltage frequency converter
45
Over temperature frequency converter
46
Over temperature motor
47
Over current frequency converter
48
Timeout between control unit and frequency converter
49
Other error frequency converter
55
Over Speed
70
Timeout between controller and RS232 interface
99
Rotor is not allowed in this centrifuge
FALSE
Inserted rotor does not exist in the program
rotor no
Rotor is not detected
© Hermle Labortechnik GmbH
Z326_V2.16