iM80 Patient Monitor Service Manual
To disassemble the CO
Separate the front and back
assemblies
Assemble the CO
module in the reverse order.
2

8.5.5 Replacing NIBP Module

Wire connecting main board
To disassemble the NIBP module:
Separate module bracket assembly
as described in section 8.3
Assemble the NIBP module in the reverse order.
module:
2
Disconnect
connecting CO
Disconnect
connecting NIBP module
wires
Unscrew
module
securing CO
2
Wire connecting main board
wires
Unscrew
securing NIBP module
- 47 -
Disassembling the Monitor
the
screws
module
2
Wire
main board
the
screws
connecting