Nibp Leakage Test; Nibp Calibration - EDAN iM80 Servise Manual

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

iM80 Patient Monitor Service Manual
within the range of ±8mmHg.

4.2.4 NIBP Leakage Test

This test checks leakage of the airway. See Figure 4-1 for details about tools required.
1.
Connect the cuff securely with the socket for NIBP air hole.
2.
Wrap the cuff around the cylinder with an appropriate size.
3.
Access Menu > Maintenance > User Maintain by inputting the password ABC. Start a
leakage test by selecting NIBP Maintain > Leak Test.
The system will automatically inflate the pneumatic system to 180 mmHg. After 20 seconds, the
system will automatically open the deflating valve, which marks the completion of a pneumatic
measurement.
If the prompt of Leak Test OK appears, it indicates that the airway is in good situation and no air
leaks exist. However if the alarm information of NIBP Cuff Leak appears, it indicates that the
airway may have air leaks. In this case, the user should check for loose connection. After
confirming secure connections, the user should re-perform the pneumatic test. If the failure
prompt still appears, please contact the manufacturer for repair.

4.2.5 NIBP Calibration

NIBP calibration must be performed by professional personnel authorized by EDAN. Access
Menu > Maintenance > User Maintain by inputting the password ABC.
NOTE:
NIBP calibration is for checking the measurement accuracy and cannot change the
measurement results.
Tools required:
• T-fitting;
• NIBP extension tubes;
Figure 4-1 Diagram for NIBP Leakage Test
- 13 -
Test and Maintenance