EDAN iM80 Servise Manual page 11

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

iM80 Patient Monitor Service Manual
1 The monitor must be serviced only by authorized and qualified personnel. EDAN does
not assume any responsibility for damage or injury if modifications or repairs are
carried out by unauthorized personnel.
2 Use and replace the substitutive parts provided or recommended by EDAN only.
3 The service personnel must be familiar with the operation of this monitor. Refer to
Patient Monitor User Manual for details.
4 Perform periodic safety test to ensure patient safety. Safety test should include
leakage current measurement and insulation testing.
5 Disconnect the monitor from power before replacing the fuses which are with the
identical specifications.
6 SHOCK HAZARD – Do not remove the top panel cover during operation or while
power is on. The unit cover must be removed only by authorized service personnel.
7 SHOCK HAZARD – Do not attempt to connect or disconnect the power cord with wet
hands. Make sure that your hands are clean and dry before touching the power cord.
8 Accessory equipment connected to the analog and digital interface must be certified
according to the respective IEC/ EN standards (e.g. IEC/ EN 60950 for data
processing equipment and IEC/ EN 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the valid version of the system standard IEC/ EN
60601-1-1. Anybody who connects additional equipment to the signal input connector
or signal output connector to configure a medical system must ensure that the system
complies with the requirements of the valid version of the system standard IEC/ EN
6060-1-1-1. If you have any question, please consult our technical service department
or your local distributor.
9 Do not remove the battery while AC power is on.
10 Do not directly connect the battery to an electric outlet.
11 Do not directly solder the lead wire and the batter terminal.
1 The device is designed for continuous operation. Avoid splashing water over the
device.
2 Do not operate the device when it is damp or wet. Avoid using the device immediately
after relocating it from a cold environment to a warm and humid environment.
WARNING
CAUTION
- 5 -
Safety Guidance

Hide quick links:

Advertisement

Table of Contents
loading