Vital Signs Monitor Service Manual
○
1
Screws
○
iM3 AC cable
2
Separate the front
and back assembly.
Assemble the power module in the reversed order. Connect the wires and fix the main unit.
8.4.2 Replacing SunTech Module
○
Screws
1
○
2
Air tube
Separate the front
and back assembly
Assemble SunTech module in the reversed order. Connect the wires and air tube, and then fix the
main unit.
Disconnect
power
module linking wire
○
1
○
3
Remove module
Unscrew the screws
wires and air
securing
tube
module
- 46 -
Disassembling the Monitor
iM3 Wires linking parameter
○
3
board and AC&DC module
Unscrew
the
screws securing
power module
○
2
iM3 Wires linking parameter
○
3
board and SunTech module
SunTech
Replace power
module.
Replace SunTech
module