DETAILED COMMAND LIST

The four letter mnemonic in each command sequence specifies the command. The rest of the sequence
consists of parameters. Multiple parameters are separated by commas. Parameters shown in { } are optional
or may be queried while those not in { } are required. Commands that may be queried have a question mark
in parentheses (?) after the mnemonic. Commands that may ONLY be queried have a ? after the mnemonic.
Commands that MAY NOT be queried have no ?. Do not send ( ) or { } as part of the command.
The variables are defined as follows.
i, j, k, l, m
x, y, z
f
s
All numeric variables may be expressed in integer, floating point or exponential formats ( i.e., the number five
can be either 5, 5.0, or .5E1). Strings are sent as a sequence of ASCII characters.
Remember!
All responses are directed only to the selected output interface!
Use the OUTX command to select the correct interface at the beginning of every program.
integers
real numbers
frequency
string
Remote Programming
5-3