Abridged Command List - Stanford Research Systems SR530 Manual

Lock-in amplifier
Table of Contents

Advertisement

Abridged Command List

AX
Auto offset X
AY
Auto offset Y
AR
Auto offset R
AP
Auto phase
B
Return Bandpass Filter Status
B0
Take out the Bandpass Filter
B1
Put in the Bandpass Filter
C
Return the Reference LCD Status
C0
Display the Reference Frequency
C1
Display the Reference Phase Shift
D
Return Dynamic Reserve Setting
D0
Set DR to LOW range
D1
Set DR to NORM range
D2
Set DR to HIGH range
En
Return Channel n (1 or 2) Expand
Status
En,0
Turn Channel n Expand off
En,1
Turn Channel n Expand on
F
Return the Reference Frequency
G
Return the Sensitivity Setting
G1
Select 10 nV Full-Scale
...
(G1-G3 with SRS preamp only)
G24
Select 500 mV Full-Scale
H
Return Preamp Status (1=installed)
I
Return the Remote/Local Status
I0
Select Local: Front panel active
I1
Select Remote: Front panel inactive
I2
Select Remote with full lock-out
J
Set RS232 End-of-Record to <cr>
Jn,m,o,p
Set End-of-record to n,m,o,p
K1
Simulates Key-press of button #1
...
(see un-abridged command list)
K32
Simulates Key-press of button #32
L1
Return Status of Line Notch Filter
L1,0
Remove Line Notch Filter
L1,1
Insert Line Notch Filter
L2
Return Status of 2XLine Filter
L2,0
Remove 2XLine Notch Filter
L2,1
Insert 2XLine Notch Filter
M
Return the f/2f Status
M0
Set reference mode to f
M1
Set reference mode to 2f
N
Return the ENBW setting
N0
Select 1 Hz ENBW
N1
Select 10 Hz ENBW
OX
Return X Offset Status
OX 0
Turn off X Offset
OX 1,v
Turn on X Offset, v = offset
OY
Return Y Offset Status
OY 0
Turn off Y Offset
OY 1,v
Turn on Y Offset, v = offset
OR
Return R Offset Status
OR 0
Turn off R Offset
OR 1,v
Turn on R Offset, v = offset
P
Return the Phase Setting
Pv
Set the Phase to v. Abs(v) <999 deg
Q1
Return the Channel 1 output
Q2
Return the Channel 2 output
QX
Return the X Output
QY
Return the Y Output
R
Return the trigger mode
R0
Set the trigger for rising edge
R1
Set the trigger for + zero crossing
R2
Set the trigger for falling edge
S
Return the display status
S0
Display X and Y
S1
Display X and Y Offsets
S2
Display R and Ø
S3
Display R Offset and Ø
S4
Display X and Y noise
S5
Display X5 and X6 (ext D/A)
T1
Return pre-filter setting
T1,1
Set the pre-filter TC to 1 mS
...
T1,11
Set the pre-filter TC to 100 S
T2
Return the post-filter setting
T2,0
Remove post filter
T2,1
Set the post filter TC to 0.1 S
T2,2
Set the post filter TC to 1.0 S
V
Return the value of the SRQ mask
Vn
Set the SRQ Mask to the value n
(See the Status Byte definition)
W
Return the RS232 wait interval
Wn
Set RS232 wait interval to nX4mS
Xn
Return the voltage at the rear panel
analog port n. (n from 1 to 6)
X5,v
Set analog port 5 to voltage v
X6,v
Set analog port 6 to voltage v
Y
Return the Status Byte value
Yn
Test bit n of the Status Byte
Z
Reset to default settings and cancel
all pending commands.
7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents