SERVICE REQUESTS (SRQ)
A GPIB service request (SRQ) will be generated whenever a bit in both the Serial Poll Status byte AND Serial
Poll Enable register is set. Use SSRE to set bits in the Serial Poll Enable register. A service request is only
generated when an enabled Serial Poll Status bit becomes set (changes from 0 to 1). An enabled status bit
which becomes set and remains set will generate a single SRQ. If another service request from the same
status bit is desired, the requesting status bit must first be cleared. In the case of the ERR, LIA and ESB bits,
this means clearing the enabled bits in the ERR, LIA and ESB status bytes (by reading them). Multiple
enabled bits in these status bytes will generate a single SRQ. Another SRQ (from ERR, LIA or ESB) can only
be generated after clearing the ERR, LIA or ESB bits in the Serial Poll status byte. To clear these bits, ALL
enabled bits in the ERR, LIA or ESB status bytes must be cleared.
The controller should respond to the SRQ by performing a serial poll to read the Serial Poll status byte to
determine the requesting status bit. Bit 6 (SRQ) will be reset by the serial poll.
For example, to generate a service request when a RESRV overload occurs, bit 0 in the LIA Status Enable
register needs to be set (LIAE 0,1 command) and bit 3 in the Serial Poll Enable register must be set
(*SRE 3,1 command). When a reserve overload occurs, bit 0 in the LIA Status byte is set. Since bit 0 in the
LIA Status byte AND Enable register is set, this ALSO sets bit 3 (LIA) in the Serial Poll Status byte. SInce bit
3 in the Serial Poll Status byte AND Enable register is set, an SRQ is generated. Bit 6 (SRQ) in the Serial Poll
Status byte is set. Further RESRV overloads will not generate another SRQ until the RESRV overload status
bit is cleared. The RESRV status bit is cleared by reading the LIA Status byte (with LIAS?). Presumably, the
controller is alerted to the overload via the SRQ, performs a serial poll to clear the SRQ, does something to
try to remedy the situation (change gain, experimental parameters, etc.) and then clears the RESRV status bit
by reading the LIA status register. A subsequent RESRV overload will then generate another SRQ.
STANDARD EVENT
STATUS BYTE
The bits in this register remain set until cleared by reading them or by the jCLS command.
bit
name
0
INP
1
Unused
2
QRY
3
Unused
4
EXE
5
CMD
6
URQ
7
PON
5-24
usage
Set on input queue overflow (too many
commands received at once, queues cleared).
Set on output queue overflow (too many
responses waiting to be transmitted, queues
cleared).
Set when a command can not execute correctly
or a parameter is out of range.
Set when an illegal command is received.
Set by any key press or knob rotation.
Set by power-on.