Enabling/Disabling the AUTO-HOLD Mode:
■ Enter the menu by ►. HOLd blinks.
■ Press ▲ or ▼ to switch the AUTO-HOLD mode to
active or inactive.
11
HOLd
- AUTO-HOLD Mode Disabled: Only the
- AUTO-HOLD Mode Enabled: The
■ Validate and return to the previous menu by ►.
Displaying the version number
■ Enter the menu by ►.
12
VER
■ The version number is displayed.
■ Return to the previous menu by ►.
Reserved for factory use - not used for normal
13
CAL
operation

4.5 Calibration Loop

The provided 5Ω calibration test loop (Figure 4-2) lets you perform a quick
check of the instrument's accuracy. To do this, turn ON the instrument and wait
until it completes its startup calibration cycle. Then clamp the instrument around
the loop; the measurement should read between 4.8Ω and 5.2Ω. If not, clean
the instrument and repeat the measurement. If you are still unable to obtain a
measurement between 4.8Ω and 5.2Ω, contact AEMC.
20
www.
Function
icon is displayed.
icons are displayed.
Figure 4-2
Clamp-on Ground Resistance Tester Model 6416