4.2
WARNING: Make sure that no terminals are connected and that the
switch is in the OFF position before opening the back of the instrument.
To replace the fuse, proceed as follows:
• Set the switch to OFF.
• Disconnect the measurement leads from the input terminals.
• Using a screwdriver, unscrew the three screws of the battery com-
• Remove the blown fuse using a screwdriver.
• Insert a new fuse (HRC FF 0.5A–600V–6.3x32), then screw the
4.3
WARNING: Make sure that no terminals are connected and the range
switch is in the OFF position.
• Clean the body of the instrument with a cloth lightly moistened
• Wipe clean with a damp cloth.
• Dry with a dry cloth or forced air.
• Do not use alcohol, solvents or hydrocarbons.
24
0.561.8187

Fuse Replacement

partment cover located on the back of the housing.
cover back onto the housing.

Cleaning

with soapy water.