3.6.1
To set the compensation of the resistance of the measurement leads, pro-
ceed as follows:
Short-circuit the black and red leads connected to the meter.
Hold the ZERO/
lowest value. The
Release the ZERO/
stored.
NOTE: Valid in V
3.7
The Model 6527 generates a DC voltage between the V-Ω and COM
terminals. It then measures the current across these two terminals and
from it calculates the value of R=V/I.
Turn the range switch to kΩ. The test voltage is generated without
pressing the test button.
Connect the black test lead to the (COM) terminal and the red one to
the (V-Ω) terminal.
Connect the test probe tips to the sample under test.
22
0.561.8187

Lead Resistance Compensation

button down until the display unit indicates the
, V
, Ω and kΩ positions only
AC
DC

Resistance Measurements

www.
symbol will display.
button. The correction value is displayed and
Figure 3-12
.com