Warning: Tests are to be carried out on non-energized circuits only!
The Model 6527 generates a 200mA
terminals. It then measures the voltage across the two terminals and from
it deduces the value of R = V/I.
Turn the function switch to
Connect the black test lead to the (COM) terminal and the red one to
the (V-Ω) terminal.
Connect to the sample under test.
The megohmmeter will continue generating a test current (200mA)
until it is switched out of the continuity function. Turn the megohmme-
ter OFF when finished and before touching the lead ends.
When there are inductive or capacitive elements in the tested sample,
it may be useful to invert the polarity by switching the leads around
on the sample. Take both measurements and use their average as a
measurement result.
Megohmmeter Model 6527
0.561.8187

Continuity Measurements