Specifications; Reference Conditions - AEMC 6527 User Manual

Megohmmeter
Table of Contents

Advertisement

5.1
5.2
ELECTRICAL
INSULATION RESISTANCE (250V)
Range
Resolution
Accuracy
Test Current
INSULATION RESISTANCE (500V)
Range
Resolution
Accuracy
Test Current
INSULATION RESISTANCE (1000V)
Range
Resolution
Accuracy
Test Current
AC VOLTAGE TRMS
Range
Resolution
Accuracy
Input Impedance
DC VOLTAGE
Range
Resolution
Accuracy
Input Impedance
Megohmmeter Model 6527
0.561.8187

SPECIFICATIONS

Reference Conditions

Influence Quantity
Temperature
Relative Humidity
Specifications
0.001M
2% ± 10cts
0.001M
2% ± 10cts
0.001M
3% ± 10cts
www.
CHAPTER 5
4M
40M
0.01M
2% ± 10cts
1mA test current into a 250kΩ load
4M
40M
0.01M
2% ± 10cts
1mA test current into a 500kΩ load
4M
40M
0.01M
2% ± 10cts
1mA test current into a 1MΩ load
600V
10MΩ (40/400Hz)
600V
.com
Reference Values
23°C ± 2°K
45 to 75%
400M
0.1M
3% ± 5cts
400M
0.1M
2% ± 5cts
400M
0.1M
3% ± 5cts
1000V
1V
1.2% ± 10cts
1000V
1V
0.8% ± 3cts
10MΩ
information@itm.
4G
1M
4% ± 5cts
4G
1M
4% ± 5cts
4G
1M
4% ± 5cts
(1)
(1)
25

Advertisement