ImageXpress Micro High Content Imaging System Options User Guide
18
Loading the Sample Plate
You must use the live-cell sealing ring to maintain the CO
sample plate. The sealing ring helps to contain the inputted air directly
over the cells and maintains the proper CO
WARNING! BIOHAZARD! Wear gloves when handling sample plates.
Note: To ensure that it is at the proper temperature, before you load
the sample plate, make sure that the live cell sealing ring is in the system
or in the incubator.
To load the sample plate
1. In the MetaXpress software, click Screening > Plate acquisition
and Control > Eject Plate to open the top door of the instrument.
2. Insert the sample plate into the stage and then remove the lid
from the plate. If there are concerns about contamination, a
breathable seal can be used over the top of the plate.
flow to the
2
and temperature levels.
2