Items to Be Provided by the Customer

The customer must provide the following items for the Environment
Control option installation:
A tank of pre-mixed 5% CO
The regulator, fittings, and tubing required to deliver 15 to 20 PSI
carbon dioxide from the tank to the Environmental Control
Option Controller or the Systems Power & Options Controller.
Microwell plates with standard ANSI height. Most 96-well and
384-well plates are standard height. The standard height for
these plates is: 14.35 mm ± 0.25 mm (0.5650 inches ± 0.0098
inches).
Deionized water to maintain humidity.
An FSE (Field Service Engineer) installs the ImageXpress Environment
Control option. After installation, environmental control can be set up for
experiments.

Setting the Temperature

The temperature controller is calibrated before the instrument ships
from the factory. Use the external temperature controller to set the
temperature that you want the environmental enclosure to maintain.
and 95% air.