Quick Troubleshooting - Kodak 2100 User Manual

Intraoral x-ray system
Hide thumbs Also See for 2100:
Table of Contents

Advertisement

Chapter 6
TROUBLESHOOTING
Occasionally, malfunctions can occur during use in the event of an incorrect action or failure. The
quick troubleshooting, the Information "Ixx" and Error message "Exx" guide you through the actions
you need to take to correct the malfunction.
IMPORTANT
If the malfunction persists or more serious conditions occur, contact your
representative and stop the equipment.
When you call your representative, you must have the following information ready:
Model Number: KODAK 2100
Serial Number (on the labels)
Error Code Number: Exx.

Quick Troubleshooting

Quick troubleshooting guides you through the actions you need to take to correct the
malfunctions.
The following table lists the malfunctions and the actions to take.
Malfunction
Nothing lights up
Control unit doesn't light up.
No X-ray emission
The X-ray emission works, but
the exposure is too light or
completely white.
Possible Cause
The unit is disconnected.
Fuse F1 burnt out or is
defective.
The circuit breaker is OFF. Turn ON the circuit breaker
The control unit is
disconnected.
Fuse F1 burnt out or is
defective.
The control unit is defective Call your representative.
The generator is cooling
The radiology control key is
defective.
The generator is not
positioned correctly.
The exposure time is too
short.
The development time is
too short.
Action
Connect the unit.
Change the fuse.
Connect the control unit.
Change the fuse.
Wait until the I01message is no longer
displayed.
Call your representative.
Adjust the generator position
Increase the exposure time.
Increase the development time (See the
manufacturer's instructions).
KODAK 2100 Intraoral X-Ray System User Guide (SM700)_Ed04
6–1