General Overview
2–2
KODAK 2100 SYSTEM OVERVIEW
•
A wall framework which contains the:
•
Main powerboard
•
ON/OFF switch which contains a built-in LED.
•
A control timer unit which:
•
Performs exposure time selection and displays parameters (exposure time and
emitted dose).
•
Performs a microprocessor self-test at every unit activation.
•
Displays alarms in case of incorrect operation.
•
Includes two exposure time modes. The digital mode corresponds to the range of
shortest exposure times that fit those needed for digital sensors.
•
An extension arm and a scissor arm which is equipped with springs that ensure arm
stability.
The following figure illustrates the extension arm and the scissor arm.
Figure 2–1
KODAK 2100 Intraoral X-ray System - Side View
R
Table 2–1
Extension arm types
Extension
Short
Standard
Long
•
A rectangular collimator. Its various sizes adapt to films and RVG sensors. It is
recommended to use such a collimator if possible, in order to limit the radiation absorbed
by patients.
A
R
47.0 cm
(18.5 in.)
64.8 cm
(25.5 in.)
82.5 cm
(32.5 in.)
Span A
170.0 cm
(67 in.)
188.0 cm
(74 in.)
205.0 cm
(80.69 in.)