Tunstall Vi User Manual page 8

Personal emergency response system
Hide thumbs Also See for Vi:
Table of Contents

Advertisement

Step 2 – Initial System Test
Press the button on the Help Activator. The two-way voice Home Console Unit
will beep and transmit a signal to the Response Center. A Response Center
operator will speak with you over the Home Console Unit to activate the
service, verify personal and responder information and ensure the system is
working properly. If you have more than one Help Activator, please be sure to
send a signal to the Response Center by pressing the button of each
activator.
IMPORTANT NOTE: The operation of your system is dependent upon compatible
phone service. A change in telephone service after proper installation may render the
system inoperative. If the phone service is changed or modified, the subscriber must
verify the system compatibility by successfully activating the system with the Response
Center. If the signal is not received by the Response Center, the phone service may
not be compatible and the system will not operate as intended.
Step 3 – Help Activator Range Test
The range of all programmed help activators can be tested without calling the
Response Center. The range of help activator(s) should be tested upon
installation or relocation of the Home Console Unit to confirm coverage
around your home.
To put the Home Console Unit into walk/range test mode, follow these simple
steps;
1.
Press and hold the GREEN RESET button on the home console unit
until it beeps. After releasing the button, the unit will announce
"PROGRAMMING MODE".
2.
Test the help activator by pressing the button. If you are in range, the
Home Console Unit will announce "PERSONAL TRIGGER" or "FALL
DETECTOR" depending on the help activator supplier to you. Test in
different locations of your home. If you do not hear "PERSONAL
TRIGGER" after pressing the help activator, you are out of range. Try
to find a more central location for the Home Console Unit to allow for
improved activator range coverage.
3.
The Home Console Unit will automatically exit range test mode after
2 minutes. To manually exit the range test mode, press the GREEN
RESET button.
IMPORTANT NOTE: It is important to test the help activator in all the areas of your
home, including the bathroom, basement and garage, as environmental conditions
such as furnishings or building structure may affect the range. An emergency call will
NOT be initiated if the help activator is pressed and it is out of range of the Home
Console Unit.