Menu 5 – Not Worn Period ....................... 12 Menu 6 – Sleep Mode (for transportation) ................12 Programming the iVi to a Lifeline home unit .................. 13 Using the iVi to answer an incoming call ..................13 Using with the pendant test reminder (Lifeline Vi/Vi+ only) ............. 13 Using Not Worn Alerts (Lifeline Vi/Vi+ only) ................
LED. The user then has 10 seconds to cancel the alert by pressing the cancel button. The iVi can send an alert to the monitoring centre if it hasn’t been worn for a period of time. This period of time is configurable (3, 5 or 7 days). This feature is turned off by default and only turned on in advanced mode.
For details on how each of the wearing options is fitted, refer to the User guide that is included in the iVi packaging. Alert Types The following table explains what alarms will be presented at a PNC and what caused that particular alarm.
Connect, CareAssist, Lifeline 400, 4000+, Connect, Connect +, Vi and Vi+? Is the user capable of using the cancel button correctly? • Once the above points are known, it is possible to program the iVi correctly and then program the iVi to the Lifeline. Identify...
Waist Worn performance. (users 5ft 10in or above) Cancel button Enabled The user can cancel an automated fall alert when the iVi is beeping. Not worn period 7 days The Not Worn alert is disabled...
Press and hold the Help button and then the Cancel key • and hold both for 3 seconds. After more than 3 seconds release the Alarm key then release the Cancel key (iVi • beeps and the green status LED lights up) The iVi is now in programming mode •...
LED. A press of the Help button within each menu will cycle between the options within that menu and is indicated by different tones played by the iVi. The settings are automatically saved when the programming menu is exited (after 20 seconds of no button presses).
Quick Programming Guide Menu 1 - Adjustable Sensitivity Green LED on Option Setting Indicator Tone Note High High Tone Very inactive user susceptible to falls Medium Medium Tone Less active user Low Tone Default setting Menu 2 – Equipment Type...
Normal Mode The normal mode is for use with all Tunstall home units and radio systems. It prevents the ‘Activation Cancelled’ and ‘Not Worn’ alerts from being sent by iVi as these alerts are not supported by all systems. Advanced Mode Must only be used with Lifeline Vi/Vi+ and when ‘Activation...
Menu 3 - Wearing Option Orange LED on The iVi can be worn in multiple positions. Menu option 1 and 2 is selected depending upon the height of the user and the wearing position of the iVi. Users below the minimum height threshold may still wear the device but they may experience a lower performance.
Pressing and releasing the Help button puts the iVi into sleep mode and turns off the LEDs. If you don’t want to enter sleep mode, press the cancel button to move to Menu 1 or press no further buttons and after 20 seconds the iVi will exit the programming menu.
Using the iVi to answer an incoming call The iVi, like a standard pendant, can be used to answer an incoming call by simply pressing the alarm button when the connected telephone or Lifeline is ringing. Once the iVi is programmed to the Lifeline Vi, open PC Connect then click on ‘Radio Triggers &...
Using Not Worn Alerts (Lifeline Vi/Vi+ only) The iVi can also assess whether it is being worn. The period before this raises an alarm is configurable (3, 5 or 7 days). This feature is turned off by default and only turned on in advanced mode.
In the unlikely event that the iVi signals an error or 3 ALB alerts are received within 30 minutes at the monitoring centre, the iVi must be replaced immediately and returned to Tunstall via the normal product returns process.