Reticle - Sokkia CX-101 Service Manual

Compact x-ellence station cx series
Hide thumbs Also See for CX-101:
Table of Contents

Advertisement

35. CHECKS AND ADJUSTMENTS
3. Wait until a beep sounds and sight
the reference point in Face 2.
Rotate the instrument 180°. Sight
the same point in Face 2 and press
[OK].
4. Set the constant.
Press [YES] to set the constant.
• Press [NO] to discard the data
and return to <Collimation>.

35.4 Reticle

PROCEDURE Check 1: Perpendicularity of the reticle to the
1. Carefully level the instrument.
2. Align a clearly visible target (the
edge of a roof for example) on
point A of the reticle line.
3. Use the telescope fine motion
screw to align the target to point B
on a vertical line.
If the target moves parallel to the
vertical line, adjustment is
unnecessary. If its movement
deviates from the vertical line,
have our service representative
adjust it.
276
horizontal axis

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cx-102Cx-103Cx-105Cx-107

Table of Contents