Calibrating High Temperature PINIP™ Probes
Special thickness standards are supplied with
F1 2 High Temperature PINIP™ Probes - see
page 44. These thickness standards should be
used in place of the calibration standard in Step 1
of "Calibration adjustment procedure" on page 27.
1. Place appropriate thickness standard over end
of PINIP™ probe.
2. Press CAL softkey.
3. Place probe on hot
reading.
4. Lift probe and then replace on hot surface to
take second reading.
5. Repeat as necessary until reading is stable.
6. Press SET to accept value.
7. Remove thickness standard from end of
PINIP™ probe.
8. Proceed with Step 2 - see page 28.
e.
The temperature of the surface used for calibration
should be equal to the temperature of the substrate
being measured.
e
surface and take a

6 STATISTICS

The Elcometer 456
3
(STATS) which calculates and displays a statistical
analysis of readings as they are taken. The
statistical calculations are also applied to the
readings stored within a single batch in memory.
When
EXTENDED
(MENU/EXTENDED MENU/SEL), press STATS
softkey to access STATS MENU (Figure 39).
Figure 39. Stats menu
The statistical values available are:
Number of Readings
Mean
Standard Deviation
Coefficient of Variation
Highest Reading
Lowest Reading
Top has a Statistics feature
MENU
is
R
active
31