How the Oscilloscope Works
System Bandwidth
Table 1-2 shows characteristics of recommended probes for the oscilloscope.
The bandwidth shown in the table is the overall system bandwidth when used
with the listed plug-in and the mainframe.
Table 1-2
HP 54701A Active Probe
Input resistance
Input capacitance
Division Ratio
Offset range
When used with
HP 54711A
HP 54712A
HP 54713B
HP 54721A
HP 54006A Passive 50-Ω Divider Probe
Input resistance
Input capacitance
Division Ratio
When used with
HP 54711A
HP 54712A
HP 54721A
HP 10430A 1-MΩ Passive Probe
Input resistance
Input capacitance
Division Ratio
When used with
HP 54713B
1–32
100 kΩ
≤0.6 pF
10:1
±50 Vdc
Bandwidth
Maximum sensitivity
1.3 GHz
20 mV/div
1 GHz
10 mV/div
500 MHz
10 mV/div
1 GHz
10 mV/div
500 Ω (10:1) or 1000 Ω (20:1)
Typically 250 fF
10:1 or 20:1
Bandwidth
Maximum sensitivity
1.5 GHz
20 mV/div (10:1) or 40 mV (20:1)
1.1 GHz
10 mV/div (10:1) or 20 mV (20:1)
1.1 GHz
10 mV/div 10:1) or 20 mV (20:1)
1 MΩ
≤6.5 pF
10:1
Bandwidth
Maximum sensitivity
500 MHz
10 mV/div