Limit Test Menu
Fail Action
The following events reset all the limit test results, which starts a new
summary file.
Changing any of the thresholds in the define measure menu.
Changing the time base window, scale, position, reference, or resolution
(number of points)
Changing the trigger definition.
Changing the sampling mode between real time and equivalent time.
Changing the filter bandwidth, if measuring filtered data.
Clearing the display.
The following events reset individual limit test results. If one of the following
events also happens to reset all of the individual limit test results, then a new
summary file is also started.
Selecting a measurement.
Turning on the display of the signal under measurement.
Changing the channel state, scale, or offset for the measurement on that
channel.
Redefining a waveform math function, if the measurement is calculated on
that function.
Storing to memory, for the measurements calculated on that memory.
Change the upper or lower limits.
Change the fail when mode.
The store summary choices are off, printer, or disk.
Off Off means that the summary data is not printed or stored.
Printer Printer sends the summary to the destination determined by
the Print selections. The destination can be one of several printers. The
printer selection gives you a hardcopy of the summary information.
See also
Setup print menu for a description of the selections.