Table 1-1 shows the common characteristics for the channel plug-ins.
Table 1-1
Characteristic
HP 54711A
Sample Rate
2 GSa/s
Bandwidth
in A models
1.5-GHz
in D models
2.0-GHz
Channels
1
Memory depth
in A models
16K
in D models
64K
Trigger
Ext. (2.5-GHz)
Logic trigger
No
Max. Sensitivity
20 mV/div
(with software
2 mV/div
expansion)
Input resistance
50 Ω
Slots used
1
*The standard 1-2-5 sequence in the 54722A plug-in, which is selected by the
mainframe's front-panel knob, does not correspond exactly to the attenuator ratios in
the hardware. Refer to the HP 54722A Attenuator Plug-in User's Reference for more
information about the attenuator ranges.
Use the HP 54711A plug-in when:
The signal is repetitive, and you are using the equivalent-time sampling
mode.
You need high equivalent-time bandwidth.
You can use external triggering, and you need high trigger bandwidth.
You can give up logic triggering.
Use the HP 54712A plug-in when:
The signal is repetitive, and you are using the equivalent-time sampling
mode.
The signal happens once or infrequently, you are using the real-time
sampling mode, and the input signal does not contain any frequency
components above 500 MHz.
You need internal triggering.
1–18
HP 54712A
HP 54713B
2 GSa/s
2 GSa/s
1.1-GHz
500-MHz
1.1-GHz
500-MHz
1
1
16K
16K
64K
64K
Internal
Internal
Yes
Yes
10 mV/div
7 mV/div
1 mV/div
1 mV/div
50 Ω
1 M Ω/50 Ω
1
1
HP 54714A
HP 54721A
2 GSa/s
4 GSa/s
400-MHz
1.1-GHz
400-MHz
1.1-GHz
2
1
8K/chan
32K
32K/chan
128K
Internal
Internal/Ext.
Yes
Yes
7 mV/div
10 mV/div
1 mV/div
1 mV/div
1 M Ω/50 Ω
50 Ω
1
2
HP 54722A
8 GSa/s
1.5-GHz
2.0-GHz
1
64K
256K
Ext. (2.5-GHz)
Yes
80 mV/div *
8 mV/div
50 Ω
4