Table of Contents

Advertisement

Mask Menu

Test

Test

The Test softkey starts and stops the mask test. Off stops the mask test. If
the histogram and color graded display features are also turned off, then
turning off the mask test also disables the data base. Off allows you to
construct a mask and configure the test options before actually starting the
mask test.
On starts the mask test. The mask test continues to run until the test is
turned off, the Stop/Single hardkey is pressed, power is cycled, or the
conditions set by the run until softkey are met.
Also, if the data base for the mask test is not already built (histograms and
color graded display turned off), on starts building the data base. If the data
base is already built, on automatically updates the mask test results as if
mask testing had been on all the time. You will know how many waveforms
and samples were taken, and how many samples failed, but not how many
waveforms failed.
Pressing the Clear display key clears the data base.