Specifications

The following are specifications used to test the HP 54700-Series
mainframes. Specifications are valid after a 20 minute warm-up period.
Time base
Time-Interval Measurement Accuracy
Real Time ±[(0.2)(sample interval) + 0.007% of delta-time marker
reading].
Equivalent Time (16 averages) ±(30 ps + 0.007% of delta time
marker reading)
Front Panel
dc Output
Calibrator
Adjustable output range −2.5 V to +2.5 V when terminated into 50
Ω.
Output Delta Voltage Accuracy ±(0.2% of delta voltage output)
1) Time interval measurement accuracy includes error sources such as time base
inaccuracy, aperture uncertainty/sample clock phase jitter, reconstruction filter
uncertainty (real-time only), trigger interpolation jitter, and channel-to-channel skew.
Real-time time-interval accuracy applies for input rise times greater than 1.4
multiplied by the sample interval.
Specifications and Characteristics
Specifications