Disk Menu
File Format
YRange YRange is the voltage across eight vertical divisions of the
display. The default value is 1.60000E-001.
YOffset YOffset is the voltage at the center of the display. The default
value is 0.
Date Date is the date when the waveform was acquired. The default
value is 10 AUG 1992
Time Time is the time when the waveform was acquired. The default
value is 01:00:00
Frame Frame is the model and serial number of the mainframe which
acquired the waveform. The default value is 54720A:314A00000.
Module Module is the model and serial number of the plug-in used in
the mainframe to acquire the waveform. The default value is empty
:314A00000.
Acq mode Acquisition mode is the sampling mode used to acquire the
waveform, either real time or equivalent time. The default value is real
time.
Completion Completion represents the percent of the time buckets in
the waveform record that contain data. The number of time buckets is
equal to the number of points in the waveform record. Completion is
ignored when it is read back into the oscilloscope. The default value is 0.
X Units X units is the horizontal scaling units set in the channel menu:
unknown, Volt, second, constant, or Ampere. The default value is
unknown.
Y Units Y units is the vertical scaling units set in the channel menu:
unknown, Volt, second, constant, or Ampere. The default value is
unknown.
Max bandwidth Maximum bandwidth is an estimation of the maximum
bandwidth limit of the source signal. The default value is 0.
Min bandwidth Minimum bandwidth is an estimation of the minimum
bandwidth limit of the source signal. The default value is 0.