Preparing the Analyzer for Use
To change the fuse
The fuse is installed at the factory to match the most commonly used line
voltage in the country of destination.
1
Unplug the power cord from the analyzer.
2
Using a small screw driver, press in and turn the fuse holder cap
counter-clockwise. Remove when the fuse cap is free from the housing.
3
Pull the fuse from the fuse holder cap.
4
To reinstall, select the proper fuse and place in the fuse holder cap.
AC Line Voltage
Range
Frequency
90-140 Vrms
47-440 Hz
198-264 Vrms
47-63 Hz
5
Place the fuse holder cap in the housing and turn clockwise while pressing in.
Voltage
Select
Agilent Part
Switch
Number
115
2110-0342
230
2110-0055
Fuse
Type
8 A 250 V Normal Blow
4 A 250 V Normal Blow
7-9