onnecting the Analyzer to an External Monitor
C T
Use the three rear panel outputs (RGB) to drive an external monitor (see chapter 1 for
specic video output characteristics). Connect the three BNC analyzer outputs (RGB) to the
corresponding monitor inputs.
If you use a monochrome monitor, operate the analyzer in monochrome mode and connect
only the green (G) output to the monitor.
LCD
he analyzer can drive both its internal display and an external monitor simultaneously.
Connect the VGA output to the corresponding monitor input.
External Monitor Requirements:
VGA Compatible
640 (horizontal) x 480 (vertical) resolution
59.83 Hz vertical refresh rate
16.716 mS vertical time
31.41 kHz horizontal refresh rate
31.840
S horizontal time
75 ohm video input impedance
video analog amplitude 0.7 Vp-p
negagive true
L logic for vertical and horizontal synchronization
Setting the HP-IB Address
In remote mode, a controller communicates through the HP-IB, identifying each instrument
on the bus by its HP-IB address. Because of this, each instrument on an HP-IB
unique address (0 through 29 are available).
Factory Setting
he factory sets the analyzer to address 16. he central processing unit (CPU) reads this
address from the rmware when the instrument is rst turned on and stores it in memory.
he address changes only when the value in memory changes, either through a front panel
entry, or when you change the rmware. With a rmware change, the address again defaults
to 16.