KROHNE OPTISONIC 6400 Handbook page 73

Portable ultrasonic clamp-on flowmeter
Table of Contents

Advertisement

OPTISONIC 6400
Operating conditions
Temperature
Temperature
Temperature
Temperature
Process temperature
Ambient temperature
Storage temperature
Pipe specifications
Pipe specifications
Pipe specifications
Pipe specifications
Material
Pipewall thickness
Liner thickness
Media properties
Media properties
Media properties
Media properties
Physical condition
Viscosity
Permissible gas content
(volume)
Permissible solid content
(volume)
Recommended flow velocity
Installation condititions
Measurement configuration
Inlet run
Outlet run
Dimensions and weights
Materials
Sensor
Converter
Trunk on wheels
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
Standard version: -40...+120°C / -40...+248°F
Sensor: -40...+70°C / -40...+158°F
Signal converter: -20...+55°C / -4...+131°F
(Humidity: 5...80%, non condensing).
-30...+80°C / -22...+176°F
(Humidity: 5...80%, non condensing).
Metal, plastic, ceramic, asbestos cement, internal / external coated
pipes (coatings and liners fully bonded to pipe wall)
< 200 mm / 7.87"
< 20 mm / 0.79"
Liquids
< 100 cSt (general guideline)
For detailed information please contact your local representative.
≤ 2%
≤ 5%
0.5...20 m/s
Single pipe, single path
Single pipe, dual path
Dual pipe, dual path
≥ 10 DN straight length
≥ 5 DN straight length
Dimensions and weights
For information refer to
Anodized aluminum (rail)
Polyamide PA12, covered with TPE soft touch layer on the sides
Polypropylene
www.krohne.com
TECHNICAL DATA
on page 77.
7
73

Hide quick links: