OPTISONIC 6400
Z mode; 2 rails (DN400...1500)
Measure the outer diameter of the pipe with the measuring band. Install the DOWN rail at the
opposite location of the pipe.
• Mount the DOWN rail in such a way that the transducer is at the marked location.
Figure 3-10: Mounting second rail in Z mode (1 traverse) using a reference point
1 Measure the distance between the transducer of the UP rail and the reference point.
2 Add the advised distance to determine the position of the second transducer.
Figure 3-11: Mounting second rail in Z mode (1 traverse) without using a reference point
1 Advised distance 1 = 2 + 3
W mode; 1 rail (DN15...150)
Follow the instructions as given in "V mode; 1 rail
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
V mode; 1 rail".
V mode; 1 rail
V mode; 1 rail
www.krohne.com
INSTALLATION