Slide Scanner Repair Manual
Notes:
•
If the slide scanner is properly focused, the oscilloscope displays the
waveform signal (Figure 3-12) of the focus and magnification target for one
color channel. Amplitude of all line pair sets should all be uniform.
•
If the slide scanner is not properly focused, the oscilloscope displays the
waveform signal (Figure 3-12 of the focus and magnification target for one color
channel. Amplitude of all line pair sets are not uniform.
•
Table 3-2 lists and defines the causes of non-uniformity (incorrect focus).
Cause
Lens not adjusted properly
Slide holder assembly warped
Defective lens
Properly Focused
Figure 3-12. Both sides of waveform intensified prior to adjustment (if required)
Table 3-2. Non-uniformity causes
65
Calibration and Adjustments
Corrective Action
Adjust lens position.
Replace slide holder assembly.
Replace lens.
Improperly Focused