Polaroid SprintScan SS35 Repair Manual page 65

Slide scanner
Table of Contents

Advertisement

Slide Scanner Repair Manual
Notes:
If the slide scanner is properly focused, the oscilloscope displays the
waveform signal (Figure 3-12) of the focus and magnification target for one
color channel. Amplitude of all line pair sets should all be uniform.
If the slide scanner is not properly focused, the oscilloscope displays the
waveform signal (Figure 3-12 of the focus and magnification target for one color
channel. Amplitude of all line pair sets are not uniform.
Table 3-2 lists and defines the causes of non-uniformity (incorrect focus).
Cause
Lens not adjusted properly
Slide holder assembly warped
Defective lens
Properly Focused
Figure 3-12. Both sides of waveform intensified prior to adjustment (if required)
Table 3-2. Non-uniformity causes
65
Calibration and Adjustments
Corrective Action
Adjust lens position.
Replace slide holder assembly.
Replace lens.
Improperly Focused

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sprintscan ss35 esSprintscan ss35 plus

Table of Contents