Polaroid SprintScan SS35 Repair Manual page 112

Slide scanner
Table of Contents

Advertisement

Slide Scanner Repair Manual
Hex Code
Error Definition
00 10 00 00
(Cont'd)
00 20 00 00
Uniformity correction
detected too many bad
pixels in a row.
*
00 40 00 00
Can not find calibration
window.
00 80 00 00
Analog calibration clips
video signal in darks.
*
SprintScan PLUS only.
* *
Not Used in SprintScan PLUS.
Table 5-4. Calibration Error Messages (Cont'd)
Diagnostics and Troubleshooting
Probable Cause
Light source too dim.
Operating too hot.
Light source too dim.
Defective main controller
PC board.
Defective CCD sensor
PC board or main controller
PC board.
112
Action
Check lamp, ND filter,
and optical path for
any obstructions.
If necessary, replace
lamp.
Allow slide scanner to
cool down or replace
CCD.
Note:
If this error stills
persists after the
necessary action
is taken, replace
sensor (CCD)
PC board.
Check lamp, ND filter,
and optical path for
any obstructions.
If necessary, replace
lamp.
If video signal looks good
and problem persists,
replace main controller PC
board.
See 00 80 00 00 hex code.
Replace CCD sensor PC
board or main controller
PC board.