Before Testing - Coding Your Meter
Once it recognizes the code number, the meter
displays the code value for 2 seconds and return
to the stand-by status.
Make sure the code number matches the code on
the strip vial.
1504
• 20 •
Before Testing - Coding the Meter
• Coding for Glucose test
C20
Caution
1. If the code is not displayed after 3 seconds, pull the test strip out of the port and re-
start the procedure from the beginning.
2. If the codes on the display and that on the test vial do not match, try another new strip.
If the mismatch persists, please contact your local representative for help.
3. If the meter does not power on, pull the test strip out of the port and re-insert the test
strip.
4. Avoid testing under direct sunlight, for a more accurate test result.
5. If you apply your sample too early, error message will appear on the display(refer to
page 50 ).
1.
Firmly insert the glucose test strip into the
TM
test port of LipidPro
meter, and then the
power will automatically turn on with the
code number, temperature and time.
2. Make sure the code number on the display
of the meter correspond with that on the
test strip vial. The color coding label of the
back on the test strip automatically
indicates the code number in this system.