Page 1
LipidPro Testing System Instructions for use Testing items: • Total Cholesterol • HDL-Cholesterol • Triglyceride • Calculated LDL-Cholesterol • Blood Glucose Test CL-LP1-1A0-01 LP-CL1-1B0-01 Rev. 2008-10...
Page 2
Dear LipidPro Testing System Owner Thank you for your choosing the LipidPro Testing System. This instruction for use includes important information you need know about the LipidPro Testing System. Please read it carefully. Our goal for LipidPro Testing System is attributed to provide the best quality healthcare products coupled with superior customer service.
Page 3
How to use Instructions for use LipidPro Testing System has two kinds of test principle according to test items: This guide includes all the information you need to use LipidPro for self testing one for blood glucose and the other for lipid profile.
Page 4
7. Cleaning Your Meter and Maintenance 8. Troubleshooting 49-51 Required but not provided 9. Product Specifications 52-54 - LipidPro Total choresterol test strip - LipidPro HDL-choresterol test strip 10. Warranty - LipidPro Triglyceride test strip - LipidPro lipid profile test strip 11.
Temperature & date of test Hour & Data of test Caution Average Test Results 1. Do not use LipidPro meter in a dry environment, especially if synthetic materials are present. Synthetic clothes, carpets, etc., may cause damaging static discharges Control Solution Sysmbol in a dry environment.
Back - 7. Do not bend, cut, or alter the test strip. 3. Total Cholesterol test strip 8. LipidPro™ test strips are for single use only. Do not re-use. 9. Do not test at the following temperature condition: Front - This test strip is for Total Cholesterol.
The performance of the glucose test strips has been evaluated in laboratory and cholesterol and 70~600 mg/dL (0.80-0.86mmol/L) for Triglyceride. in clinical tests (Please refer to strip instructions for use for more details). Measurement Range: The measurement range of the LipidPro Glucose Testing is 10 to 600 mg/dL (0.6-33.3mmol/L).
Test Strip as soon as a result is displayed. How to install /replace the batteries: By pressing ▲▼, adjust year and press 1. Open the battery door on the back of the LipidPro MODE ( ) button.
LipidPro Testing System Setting Your Meter Before Testing - Coding Your Meter LipidPro has two coding systems: Lipid Profile test - each vial of test strip has RFID-tag which enables the meter to recognize its code number automatically. By pressing ▲▼, select testing...
1 second and return to the 1. Firmly insert the glucose test strip into the stand-by status. test port of LipidPro meter, and then the power will automatically turn on with the code number, temperature and time.
As a quality control, or con- trol solution, commercial human serum control can be used. It is available online at sales@infopia21.com or call at +82-31-423-6170 of Infopia Customer Service. Caution...
•After following the appropriate steps and the control solution test still falls out of solution on clean bowl range, do not perform a test. Please contact the Infopia customer service at (ex. Aluminium foil). +82-31-423-6170 or sales@infopia21.com.
Testing supplies 3. Place a drop of control solution on a clean, dry surface(ex. The lid of the test strip vial). To perform a blood test you need: • • LipidPro meter •Test strips • • Sterile lancet 4. Insert a test strip into the port firmly and press •Capillary rod...
LipidPro Testing System Collecting a Drop of Blood Testing Your Blood Performing Your Test 1. Lipid Profile test Capillary blood process When you collect the sample, please use the capillary rod provided with the test strip package. Make sure enough sample has been collected to the capillary rod.
LipidPro Testing System Testing Your Blood Quick reference of test procedure -Lipid test This is for professtional use 1) After code recognition, press mode ( ) but- venouse whole blood and serum process for Lipid Profile test ton again For venous whole blood and serum, acceptable anti-coagulants are EDTA, heparin, citrate and fluoride/oxalate, and these samples may be stored at 4℃~ 20℃...
LipidPro Testing System Quick reference of test procedure -Glucose test Test Procedure 1. Lipid test - Test for multi items 1) Insert the glucose test strip. Code number is automatically displayed. Compare with code number printed on test strip vial. If...
LipidPro Testing System Test Procedure Test Procedure 2. Test for each single item 1) Total Cholesterol test Please apply the sample to the second test area of test strip until you hear beep sound. Then, the symbol, disappears. Press mode (...
LipidPro Testing System Test Procedure Test Procedure The symbol of blood is blinking. The meter displays , blinking in order, which means it is under test. Please apply the sample to the test area of When the test is completed (in about maxi- test strip until you hear beep sound.
Page 19
LipidPro Testing System Test Procedure Test Procedure 3) Triglyceride test The meter displays , blinking in order, which means it is under test. Press mode ( ) button after a test strip is inserted and confirm if the meter recognizes the right test strip by checking out the corre- sponding symbol on its display.
LipidPro Testing System Test Procedure Glucose test strip Discard Function 3. Glucose test When the test strip is inserted, the meter displays the code number for 3 seconds. 1. After checking your test result, slide the ejector button twice forward to The symbol of blood is blinking.
The figure shown Make sure there is no cream or lotion on the test site. above displays the areas where you can test using your LipidPro Testing Thoroughly dry your hands and test site.
LipidPro Testing System Reviewing Your Results Reviewing Your Results Every test result is stored with its date and time of test. 1. Total Cholesterol Following is the memory capacity of each test item in the meter. Press mode ( ) button.
LipidPro Testing System Reviewing Your Results Reviewing Your Results 3. TG Cholesterol 5. Glucose Press mode ( ) button. Press mode ( ) button. Press mode ( ) button and the meter displays the Press ▲▼ for reviewing the rest result.
LipidPro Testing System Deleting your Results LipidPro Testing System Software You can transfer test results from the LipidPro Testing System meter to a computer. deleting individual test result You can download the software from our web site. (http://www.infopia21.com) To delete any individual test result in the memory press ▲...
A cloth dampened with water and mild detergent can be used to wipe down the tion does not work. tomer support when the problem Oxidization voltage is outside of the meter. Your LipidPro meter is a precision instrument. Please out of range, etc. persists.
LipidPro Testing System Troubleshooting Troubleshooting Message Probable cause Message Probable cause Action required Action required Please wait for the meter dis- glucose : less than 10mg/dL This message appears The test result is when blood sample is plays the icon blinking before ( 0.6mmol/L )
LipidPro Testing System Troubleshooting Product Specification Inconsistent or unexpected test results Devision Content If you continue to get unexpected results, check your system with control solu- tion. If you experience symptoms that are not consistent with your lipid and glu-...
This warranty is subject to the following exceptions and limitations: Operation Humidity 10 ~ 90% 1. Infopia Co., Ltd. shall not be required to replace any unit which are dam- aged or malfunction due to abuse, accidents, alteration, neglect, misuse, Display Type maintenance by someone other than Infopia Co., Ltd., or failure to operate...
LipidPro Testing System Declaration of Conformity Memo THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. Operation is subject to the following two conditions; (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.