Datamax I-class Maintenance Manual page 15

Hide thumbs Also See for I-class:
Table of Contents

Advertisement

Step
Action
Use the
the 00 GAIN setting and then
press the
8
Record the sensor reading as a
Label Value in a table similar
to the one shown below.
Step
Action
Press the
increment the Gain Number
then press the
record the resulting Label
9
Value.
Repeat this step for each of the
remaining Gain Numbers (01-
31).
Proceed according to your
media type:
Die-cut – peel media from
the backing and place the
backing under the Sensor
Eye Mark.
Notched – place the notch
under the Sensor Eye Mark.
10
Reflective – place the black
mark facedown under the
Sensor Eye Mark.
Use the
the 00 GAIN setting and then
press the
the sensor reading as a TOF
Value.
(continued)
Adjustments and Maintenance
Displayed Message
GAIN TRAN
Key to scroll to
*00
Key.
Or, for reflective media:
GAIN REFL
*00
Sampling Table
Gain
Label
Number
Value
00
255
01
02
...
31
Displayed Message
Key to
GAIN TRAN
*31
Key and
Or, for reflective media:
GAIN REFL
*31
GAIN TRAN
*00
Or, for reflective media:
GAIN REFL
*00
Key to scroll to
Key. Record
<yyy>
<0 - 31>
<yyy>
<0 - 31>
TOF
Difference
Value
Value
<yyy>
<0 - 31>
<yyy>
<0 - 31>
<yyy>
<0 - 31>
<yyy>
<0 - 31>
2-5
Comment
Selection is denoted by an
asterisk (*).
The sensor reading equals
the "yyy" value.
Comment
Never position a
perforation in the sensor
when recording a reading.