Chapter 4 Features and Functions

State Storage

State Storage
The 3499A/B/C provides the capability to store setups, and then recall
them to put the instrument back into a known configuration.
Information that can be stored includes:
Status of all relay channels (open or closed)
Status of digital I/O ports (input or output, flow control, etc.)
Remote interface settings (address, baud, etc.)
Additionally, you can include a stored state in a scan list as a channel.
Firmware Revision Differences
State storage behavior differs depending upon the firmware revision of
your 3499A/B/C (see page 55).
Firmware revision 1.0 must be upgraded. Please contact your nearest
Agilent Technologies Office for details.
Firmware revisions 2.0 and 3.0 allow up to 10 states to be stored.
Firmware revision 4.0 allows up to 40 stored states.
Stored states are lost if power is removed for firware revisions 2.0 or
3.0. Stored states are preserved when power is removed for firmware
revision 4.0.
Storage memory locations are numbered 01 to 10 for Firmware
revision 2.0 and 3.0. Memory locations are number 01 to 40 for
Firmware revision 4.0.
Before recalling a stored setup, the instrument verifies that all
module types and slot assignment match the setup. If a mismatch is
detected, an error ("RECALL FAILED") will occur.