Agilent Technologies 3488 User Manual page 56

Switch/control system
Table of Contents

Advertisement

Control
Firmware
Module
Version
Version
1.0
1.0
3
2.0
1.0
3.0
1.0
1
4.0
2.0
Chapter 3 System Overview
Firmware and Control Module Description
State
Plug-in module
Storage
Support
Stored setups
Must upgrade
are cleared if
hardware and
power is
firmware.
cycled.
Stored setups
Supports all
are cleared if
modules except:
power is
N2272A
cycled.
N2276A/B
N2282A
Stored setups
SCPI Mode
are cleared if
supports all
power is
modules.
cycled.
3488A Mode
supports all
modules except:
N2272A,
N2276A/B, or
N2282A.
Stored setups
SCPI Mode:
are saved.
supports all
Instrument
modules.
can be
3488A Mode
programmed
supports all
to power up in
modules except:
a set state.
N2272A,
N2276A/B, or
N2282A.
1
Upgrading to firmware revision 4.0 requires the control module be upgraded to revision 2.0.
Control module revision 1.0 does not support firmware revision 4.0.
56
Firmware
Programming
Upgrade
Modes
Needed?
Not Applicable
YES
Firmware
Upgrade to
Version 2.0
Firmware
allows
Revision 3.0
programming in
ONLY if using
both SCPI
N2272A,
Mode and
N2276A/B, or
3488A Mode.
N2282A.
Firmware
Upgrade to
Version 3.0
firmware revision
limits
4.0 ONLY if
programming to
control board
either SCPI
revision is
Mode or 3488A
upgraded to 2.0.
Mode (not both)
Firmware
NO
Version 4.0
allows
programming in
both SCPI
Mode and
3488A Mode.