To Configure the Power-on State

Firmware Rev 4.0 ONLY. To read your firmware version, see the
procedure on page 48. For more information about firmware revisions,
see "Firmware and Control Module Description" on page 55.
An instrument with Firmware REV 4.0 or later can be set to power on to
the reset state (see "Factory Default and Reset States" on page 71) or to a
state previously stored in a specified memory location. The instrument
will return to the specified state the next time it is turned on.
1. Press the Menu key. The CONFIG annunciator lights up and the
display shows the first menu level.
CARD PAIR
2. Turn the knob until "POWER ON SET" is displayed.
POWER ON SET
3. Press Enter to show the second level menu.
PWR ON RESET
4. To set the instrument power-on to the reset state, press Enter. The
instrument will return to the first level menu.
Alternately, to set the instrument power-on to a stored state: Turn
the knob until "USER SET UP" is displayed.
USER SET UP
5. Press Enter to select the stored state to use.
POWER ON 05
Chapter 2 Front-Panel Operation

To Configure the Power-on State