Error Messages; Self Test Errors - Agilent Technologies 3488 User Manual

Switch/control system
Table of Contents

Advertisement

Note
6
Note

Error Messages

In 3488A mode, you can query the instrument error queue when the
ERROR annunciator is on. The returned decimal value is the sum of the
values of the possible error conditions, as shown below:
Weighted Value
1
2
4
8
16
When the error queue is empty, the string "+00000" is returned in
response to a remote interface ERROR command. The mainframe display
will show "0000".

Self Test Errors

The errors listed below indicate failures that may occur during a self-test
(in 3488A mode).
Error Number
+1
+2
+3
+4
The string "+0" returned from a TEST command indicates that all the
tests have passed. In this case, "PASSED" displays on the front-panel of
the instrument.
136
Error Condition
Syntax Error
Execution Error which includes:
a. Parameter out of range;
b. Module type mismatch;
c. Attempt to access a nonexistent stored state
or scan list.
Hardware Trigger too fast
Logical Failure
Power Supply Failure
Description
ROM test failed.
GPIB test failed.
RS-232 test failed.
Front-panel test failed.