• When Slot 0 is specified, only Port 090 is valid.
• If a 16-bit port is specified, data will be sent with the high or most
significant byte first. The data transfer is terminated with the
3499A/B/C setting the GPIB EOI true line concurrently with the last
data byte sent.
• If a 32-bit mode (Port 06) is specified, data will be sent with the
highest byte fist, then the second highest byte and so on. The data
transfer is terminated with the 3499A/B/C setting the GPIB EOI true
line concurrently with the last data byte sent.
DIM A$[30]
IOBUFFER A$
A$ = "DBR 504,1"
TRANSFER A$ TO 709 INTR
ENTER 709 USING "%,K"; B$
DISP B$
Chapter 5 Remote Interface Reference
3488 Mode Commands
! One digital binary reading.
! Qualifiers "%,K" set PC to
concurrent EOI and last
data byte.
4