YPP6830 Explosion-Proof Pulse Input Rate/Totalizer

Specifications

Except where noted all specifications apply to operation at +25°C.

General

Display
Top
Five digits, (0-99999),
0.7" (17.8 mm) high,
7- segment,
automatic lead zero blanking.
Bottom
Seven alphanumeric characters,
0.4" (10.2 mm) high,
14-segment,
automatic lead zero blanking.
Symbols Total, grand total, battery
power/low battery, high alarm,
low alarm, SafeTouch button
sleep mode/disable, password
lock
Display
Top Display: Rate or total
Assignment
Bottom Display: Combinations of rate,
total, grand total, units, and custom tag
Backlight
White; it can be disabled/enabled in the
Advanced - System menu. The backlight
is automatically turned off below -20°C.
Alarm
Flashing display plus HI/LO (rate) or SET
Indication
(total) indicators
Display
Display may be mounted at 90°
Orientation
increments up to 270° from default
orientation.
Display
Ambient > -20°C: 1 update/second
Update Rate
Ambient < -20°C: 1 update/10 seconds
Note: Update is dependent on gate
settings.
Overrange
Display flashes
Programming
Four SafeTouch through-glass buttons
Method
when cover is installed. Four internal
pushbuttons when cover is removed.
Free PROPLUS EX software.
Recalibration
Recalibration is recommended at least
every 12 months.
Max/Min
Max/Min readings reached by the process
Display
are stored until reset by the user or until
power to the meter is turned off.
Password
Three programmable password selections
Menu Options
can be used for the following: restrict
modification of settings, prevent resetting
the total or grand total without the
password, or permanently lock out the
ability to change or reset the grand total or
any grand total related settings (making a
non-resettable grand total).
Pass: Restricts modifications of
programmed settings to require re-
entering the password to make changes.
Pass T: Restricts the reset of total to
require re-entering the password. Disables
the manual mode reset contact.
Pass GT: Restricts the reset of grand total
to require re-entering the password. May
enable a non-resettable grand total and
permanent lockout of grand total-related
settings with a specific password.
Power
9-30 VDC Powered, 2.2 W max
Options
4-20 mA Output Powered, 30 VDC max
Battery Power
9-30 VDC Powered with Battery Backup
4-20 mA Output Powered with Battery
Backup
99999
6
Battery
3.6 V Primary Lithium (Li-SOCl2),
non-rechargeable Model YPPABAT36C
Expected Service Life & Recommended
Replacement Interval
Operating
Condition
Open
collector
outputs off,
SafeTouch
buttons off,
minimal
backlight
use
<100 Hz
open
collector
outputs,
minimal
SafeTouch
button or
backlight
use
<2 kHz
open
collector
outputs,
minimal
SafeTouch
button or
backlight
use
<5 kHz
open
collector
outputs,
minimal
SafeTouch
button or
backlight
use
Backup
power only
Data Logging Up to 1024 records, recorded 4/day at
specific times or at defined time intervals.
Record contains date, time, rate, total,
grand total, and log number.
Isolation
All Models:
YPP6830-BTA:
YPP6830-APA:
Environmental Operating temperature range: -40 to 75°C
Storage temperature range: -40 to 75°C
Backlight deactivated below
temperatures ≈ -20°C
Relative humidity: 0 to 90% non-condensing
Printed circuit boards are conformally coated.
Non-Volatile
All programmed settings and total reading
Memory
are stored in non-volatile memory for a
minimum of ten years if power is lost.
Connections
Screw terminals accept 12 to 22 AWG
wire
Instruction Manual
Estimated
Suggested
Service
Replacement
Life
Interval
7.5 years
5.5 years
5.5 years
4 years
2.5 years
2 years
1.3 years
1 year
N/A
10 years
500 V opto-isolated
input-to-power/output
with isolated input
enabled
500 V input-to-output
500 V
input/power-to-output
Note: Requires
separate output supply