6. Press
DEFINE RECEIVER
CONSTANT FREQUENCY
7. Press
DONE > MULT. SRC. ON/SAVE
An
annotation should appear to the left of the graticule.
M
Other Settings
Other settings for pulsed-RF operation are as follows:
1. Press
STIMULUS MENU > POWER MENU
POWER SOURCE 1
NOTE
Set source 1 power to the highest level possible without IF overload. Power
must be set in the "normal BW" mode, then changed to the "wide BW" mode,
if desired.
POWER SOURCE 2
Check the sweep mode:
1. Press
STIMULUS MENU > STEP > SYSTEM > MORE > PULSE CONFIG > DETECTOR: WIDE
.
BW
A
annotation should appear to the left of the graticule.
W
2. Press
PULSE OUT: HIGH
3. Press
SYSTEM > MORE> SYSTEM PHASELOCK>NONE
4. Press
SYSTEM > MORE > POWER LEVELING
SOURCE1: INTERNAL
SOURCE2: INTERNAL
Agilent 85110L Test Set
.
should be 20 MHz.
.
.
should be +5 dBm.
should be +10 dBm.
.
.
.
Hardware and Instrument States
Introduction
A-5