Loading Test Images; Sending No-Burn Images To The Pacs - GE Healthcare Optima XR200amx Installation Manual

Hide thumbs Also See for Healthcare Optima XR200amx:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5336113-1EN, R
IRECTION
2.
3.
4.
5.

5.4.2.2 Loading test images

1.
2.

5.4.2.3 Sending no-burn images to the PACS

1.
2.
3.
4.
5.
10
EVISION
responsible for providing evidence of PACS monitor calibration.
At the application desktop, press IMAGE TOOLS.
Click SMPTE to load the SMPTE pattern into the Patient List.
Send the SMPTE pattern to the PACS review workstation
Verify that 5% and 95% squares are equally visible on PACS
Figure 5-2 SMPTE pattern
Load the PACS test images onto the acquisition workstation. Use either:
-
A DICOM-formatted CD-ROM with images.
-
Images that are stored on the system in Image Tools.
Verify that the patient name is GE_DX_Image_Display_Test, GEHC and the patient ID is
2000002.
In the SUIF, press the UTILITIES tab button.
Next to Network Connections, press EDIT.
Select the appropriate network host and press EDIT.
On the Preferences tab, verify that the following parameter is NOT checked:
Apply Burn-on-Send to images when sending to this network host.
In the Images list, highlight PACS Test Image #1.
Chapter 5 Calibration
O
XR200
/XR220
PTIMA
AMX
I
M
NSTALLATION
(Figure
5-2).
Page 113
AMX
ANUAL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Healthcare optima xr220amx

Table of Contents