5336113-1EN, R
IRECTION
3.5.2 Performing leakage current testing
If ground testing is required for this system, verify that the ground testing results (see
< 0.1 Ohms before conducting leakage testing.
Leakage testing must be performed if the AC power plug has been replaced or disturbed.
The leakage current must not exceed 300 uA.
Before you begin, verify these steps:
1.
2.
3.
Perform leakage testing under the five specific conditions outlined in
Condition
Table 3-4 Leakage current test conditions
10
EVISION
Complete ground testing, if required (see
page
37).
Ensure that all system covers are installed.
Obtain a Dale 601/601E test meter or equivalent.
Test meter
Power
1
Off
Normal
2
Off
Reversed
3
On
Normal
4
On
Reversed
5
On
Normal
Chapter 3 Installation
3.5.1 Performing ground resistance testing on
X-ray on
outlet
3.5.2.1 Condition 1: Power off, normal outlet,
No
No
3.5.2.3 Condition 3: Power on, normal outlet,
No
No
3.5.2.5 Condition 5: Power on, normal outlet,
Yes
O
XR200
PTIMA
I
NSTALLATION
Table
3-4:
Section reference
X-ray off
3.5.2.2 Condition 2: Power off, reversed
outlet, X-ray off
X-ray off
3.5.2.4 Condition 4: Power on, reversed
outlet, X-ray off
X-ray on
/XR220
AMX
AMX
M
ANUAL
Table
3-3) are