IRECTION
instructions to change the default internal IP Address if an IP Address conflict occurs with the
hospital network.
Note:
To change IP address or configure the snapshot tool, you need the Account and Password
information. Please consult your administrator for them.
1.
2.
3.
Note: Please consult your administrator for the Account and Password information.
4.
5.
6.
7.
8.
Note: Please consult your administrator for the Account and Password information.
xruser@magic: su
Password: #superxr
9.
10. This step applies to Optima XR220amx or Optima XR200amx with Digital Upgrade only.
11. This step applies to Optima XR220amx or Optima XR200amx with Digital Upgrade only.
12. Perform the Snapshot Tool Configuration.

4.13.1 Snapshot Tool Configuration

To ensure that the Auto-Snapshot Tool works after changing the IP Address.
1.
2.
Note: Please consult your administrator for the Account and Password information.
10
EVISION
Plug in the Service Keyboard.
Open a terminal window using CTRL+SHIFT+F5 and click on the terminal window icon.
Change users by entering the account and password at the prompt:
su
#superxr
Run the Change IP script by entering the following text at the prompt:
/magichome/xruser/engg/changeIp.sh
Select 1 of the 5 options of subnet IP addresses. Write down the IP Address you select, you
will need this in a later step.
1.
172.16.x.x
2.
172.17.x.x
3.
172.18.x.x
4.
192.168.x.x (SW LFC default)
5.
192.169.x.x
The PC will reboot automatically, select SHUTDOWN when the system presents the Degraded
mode screen.
Login to system.
Open a terminal window and become root.
At prompt, type: ifconfig | more
Confirm the br0 IP Address is XXX.XXX.3.1 (where XXX.XXX equals the address range
selected above).
To test network connectivity, perform a C-Echo test to one or more of the Network Hosts.
Verify that all connected detectors boot successfully. Verify that you can take an exposure on
all available receptors.
Verify the Detector Wireless connection is active by acquiring an image.
Open a terminal window using CTRL+SHIFT+F5 and click on the terminal window icon.
Switch to root by entering the following text at the prompt:
su
Chapter 4 Configuration
O
XR200
/XR220
PTIMA
AMX
I
M
NSTALLATION
ANUAL
Page 101
AMX

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Healthcare optima xr220amx

Table of Contents