The detector also has a backup tether that can be attached, if needed, to provide input power and
wired communication. The backup tether must be attached to the detector to register the detector
with the system, and can then be removed and stored.
3.6.1 Unpacking the detector
The detector packing is designed to protect the detector during shipping and handling. There are
three boxes inside each other to give maximum detector protection.
1.
2.
3.
4.
5.
6.
7.
10
EVISION
Locate the detector shipping box. See
Figure 3-19 Detector shipping box
Inspect the box for physical or water damage. If present, document the damage.
Remove all four clips and lift top of outer box from bottom. Instructions are printed on the
outside of the box.
Lift the plastic bag and inner box out of the outer box. See
printed on the outside of the inner box.
Figure 3-20 Inner box
Unfasten the bag straps and box straps.
Remove the box end and pull out the black inner box. See
Figure 3-21 Detector box
Using the box flaps, open the box.