3.6.2 Inspecting the detector

Visually inspect the detector for any signs of physical damage that might have occurred during
shipping. Check the corners for any signs of sharp edges or damage. Check all surfaces for any
signs of scratches or abrasions.

3.6.3 Charging the detector battery

For all customers that do not support the USA style wall plugs, locally obtain and install a computer
style power cord with a CE approved, IEC 60320-1 C13 style connector on one end and the locally
appropriate wall plug connector on the other. The cord must meet or exceed the following electrical
specification, AC 120-240V 50/60 Hz 130VA - 168VA 1.4 Amps. Do not modify the shipped US style
cord as this can present a shock hazard to the user.
The detector is delivered with two batteries and a desktop charger, which can be used to charge
the batteries when disconnected from the detector. The battery can also be charged by inserting it
into the detector, then docking the detector in the bin.
It can take up to eight hours to charge the detector battery, depending on remaining charge. The
battery LEDs (labeled with 1 in
Page 52
10
EVISION
If no damage is observed, proceed to the next step.
If damage is found, contact GE Service.
Section 3.6 Installing the digital detector
Figure
3-22) on the detector indicate remaining power.